• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2022
Event Type  malfunction  
Event Description
It was reported that the g7 straight cup inserter stuck onto a 56 g7 shell.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: 110017105- g7 finned 4 hole shell 56f- 7193292.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 01619.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product location unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No additional information on the reported event.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records; lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15003043
MDR Text Key299405566
Report Number0001825034-2022-01618
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
Patient SexFemale
-
-