• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ PLUS STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ PLUS STOPCOCK Back to Search Results
Catalog Number 394910
Device Problems Leak/Splash (1354); Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Event Description
It was reported that the cap was loose on the bd connecta¿ plus stopcock, leading to leakage.This event occurred 36 times.The following information was provided by the initial reporter, translated from japanese: the dimension of the male luer was too large to apply lock ring, causing leakage.
 
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the cap was loose on the bd connecta¿ plus stopcock, leading to leakage.This event occurred 36 times.The following information was provided by the initial reporter, translated from japanese: "the dimension of the male luer was too large to apply lock ring, causing leakage.".
 
Manufacturer Narrative
H6: investigation summary since no photos or samples displaying the reported condition of loose component - leak were available for examination, we were unable to fully investigate this incident.A device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA¿ PLUS STOPCOCK
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15008395
MDR Text Key300343345
Report Number9610847-2022-00257
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number394910
Device Lot Number1215570
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/14/2022
Initial Date FDA Received07/13/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-