Model Number 26606
Device Problems Mechanical Problem (1384); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  malfunction  
Event Description
It was reported that the stent was partially deployed and difficult to reconstrained.The 90% stenosed target lesion was located in the moderately tortuous internal carotid artery.A 10.0-24 carotid wallstent was advanced for treatment.Halfway during deployment, positioning was performed.The stent was tried to be retracted once to change the placement location, but it was noticed that it was very hard to retract the stent.So, the stent was retracted with the tip slightly extended without fully retracting it.Eventually, the device was removed.The stent was tried to be retracted outside the patient's body, and it was able to retract somehow, but it was too hard.The procedure was completed with another of same device.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15008810
MDR Text Key295864203
Report Number2124215-2022-23552
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26606
Device Catalogue Number26606
Device Lot Number0027812572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-