Model Number INTELLIVUE MULTI MEASUREMENT SERVER X2
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 05/27/2022
Event Type  Injury  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
It was reported the philips monitor became loose from the mount and fell, striking the patient¿s forehead.
 
Event Description
It was reported the philips monitor became loose from the mount and fell, striking the patient¿s forehead.The patient required sutures.Philips received a medwatch form with information on the event.There has been additional information obtained since that time.The customer noted that the mount was installed by philips, but there are no service records.The customer noted that the latch broke which caused the device to fall and hit a patient.The cause of the event was a broken latch.The customer replaced the latch to resolve the issue.The device remained on site.
 
Manufacturer Narrative
H3 other text: the customer resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MULTI MEASUREMENT SERVER X2
Type of Device
INTELLIVUE MULTI MEASUREMENT SERVER X2
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
jeanne ahearn
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15013316
MDR Text Key295865986
Report Number9610816-2022-00375
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838000261
UDI-Public00884838000261
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Risk Manager
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberINTELLIVUE MULTI MEASUREMENT SERVER X2
Device Catalogue NumberM3002A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/20/2022
Initial Date FDA Received07/14/2022
Supplement Dates Manufacturer Received10/31/2022
Supplement Dates FDA Received11/04/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-