Model Number INTELLIVUE MULTI MEASUREMENT SERVER X2 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 05/27/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
A follow-up report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported the philips monitor became loose from the mount and fell, striking the patient¿s forehead.
|
|
Event Description
|
It was reported the philips monitor became loose from the mount and fell, striking the patient¿s forehead.The patient required sutures.Philips received a medwatch form with information on the event.There has been additional information obtained since that time.The customer noted that the mount was installed by philips, but there are no service records.The customer noted that the latch broke which caused the device to fall and hit a patient.The cause of the event was a broken latch.The customer replaced the latch to resolve the issue.The device remained on site.
|
|
Manufacturer Narrative
|
H3 other text: the customer resolved the issue.
|
|