• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0702
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bradycardia (1751); Low Blood Pressure/ Hypotension (1914)
Event Date 06/24/2022
Event Type  Injury  
Manufacturer Narrative
The device was retained by hospital and not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed for a vasovagal reaction that required cardiopulmonary resuscitation (cpr) and medication.It was reported that on 6/24/2022, the patient presented with grade 3-4 mixed mitral regurgitation (mr), and a prolapse flail on the posterior medial lateral (pml).After insertion of the mitraclip steerable guide catheter (sgc) into the anatomy, the patient experienced a vasovagal reaction, resulting in hypotension and decreased heart rate, requiring cardiopulmonary resuscitation (cpr) for approximately 2 minutes.Catecholamine was also administered to stabilize the patient.There was no device malfunction.The procedure was completed with both clips implanted, reducing mr to <1.The patient left the cath lab in stable condition.No additional information was provided regarding this issue.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.The reported patient effects of hypotension and bradycardia are listed in the mitraclip g4 system instructions for use (ifu) as known possible complications associated with mitraclip procedures.Based on the information reviewed, the reported bradycardia and hypotension appeared to be related to patient conditions.The reported unexpected medical intervention and medication were result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442-4628
6517565400
MDR Report Key15020073
MDR Text Key295950042
Report Number2024168-2022-07721
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/15/2023
Device Catalogue NumberSGC0702
Device Lot Number20216R140
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/24/2022
Initial Date FDA Received07/14/2022
Supplement Dates Manufacturer Received09/09/2022
Supplement Dates FDA Received09/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STEERABLE GUIDE CATHETER
Patient Outcome(s) Required Intervention;
-
-