Device evaluation of the monitor has been completed.During the incoming functional testing, a 1hz simulated normal sinus rhythm signal was applied to the ecg electrodes, followed by a 5hz simulated treatable arrhythmia signal which verified proper performance of the detection algorithm.During the transition to the 5hz signal, the device was confirmed to properly enter into a treatment sequence which includes a verification of the tactile vibration alarm, audio messaging, and siren alarms, as well as a test of the pulse delivery circuitry.The pulse delivery circuitry test verified proper charging of the high voltage capacitors and proper delivery of five full energy 150j biphasic pulses.The functional testing confirmed proper response button functionality, ecg acquisition, detection algorithm performance, and pulse delivery functionality.There is no indication of a product malfunction.The electrode belt has been returned and the evaluation is underway.The device flag data from the last download does not indicate any device malfunction.
|
A us distributor contacted zoll to report that a patient passed away on (b)(6) 2022 while reportedly wearing the lifevest.The patient received five inappropriate shocks. the device was started up at 15:04:27 on (b)(6) 2022.At 15:51:47, an arrhythmia was detected.Ecg shows sinus bradycardia/rhythm from 30-70 bpm with cpr/motion artifact t.At 15:52:22, the patient received the first inappropriate treatment.Rhythm at the time of treatment was sinus rhythm @ 60 bpm with pvcs and pauses.Post shock rhythm was sinus bradycardia @ 50 bpm with pvcs and pauses.At 15:52:59, the patient received the second inappropriate treatment.Rhythm at the time of treatment was sinus bradycardia @ 40 bpm with pauses.Post shock rhythm was sinus bradycardia @ 40 bpm with pauses.At 15:53:27, the patient received the third inappropriate treatment.Rhythm at the time of treatment was sinus bradycardia @ 40 bpm with pauses.Post shock rhythm was sinus bradycardia @ 50 bpm with pauses.At 15:53:54, the patient received the fourth inappropriate treatment.Rhythm at the time of treatment was sinus bradycardia @ 50 bpm with pauses.Post shock rhythm was sinus bradycardia @ 50 bpm with pauses.At 15:54:22, the patient received the fifth inappropriate treatment.Rhythm at the time of treatment was sinus bradycardia @ 50 bpm with pauses.Post shock rhythm was sinus bradycardia @ 40 bpm with pauses.Oversensing of cardiac activity and cpr/motion artifact contributed to the false detection.The electrode belt was disconnected at 15:55:15 on (b)(6) 2022.The patient passed away on (b)(6) 2022.
|