• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. ARGYLE; CATHETER, PERITONEAL, LONG-TERM INDWELLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. ARGYLE; CATHETER, PERITONEAL, LONG-TERM INDWELLING Back to Search Results
Model Number 8888415604
Device Problem Connection Problem (2900)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the adapter did not connect/fit to the catheter.There was no reported patient outcome.
 
Event Description
According to the reporter, the two adapters did not connect/fit to the catheter.Nothing unusual was observed on the devices prior to use.No damage was noted on the device packaging.The box that the devices came in was intact.The devices were not repaired.Tego was not utilized.No adapter issues and no visible defects were noted.The patient had no medical intervention/treatment due to the event.It was said that the two adapters occurred on the same patient and same event date.Also, the two adapters were no longer used due to the issue and they pulled out another one (same manufacturer) to resolve the issue.There was no reported patient injury.
 
Manufacturer Narrative
Additional info: b3, b5, g1 (manufacturer name, mfr contact first name, last name, street 1, mfr city, region, postal code, email, phone number), g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the two adapters did not connect/fit (do not screw) to the catheter.Nothing unusual was observed on the devices prior to use.No damage was noted on the device packaging.The box that the devices came in was intact.The devices were not repaired.Tego was not utilized.No adapter issues and no visible defects were noted.The patient had no medical intervention/treatment due to the event.It was said that the two adapters occurred on the same patient and same event date.Also, the two adapters were no longer used due to the issue and they pulled out another one (same manufacturer) to resolve the issue.There was no impact on the patient.There was no reported patient injury.
 
Manufacturer Narrative
Additional info: b5, g3 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: d9, g2, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the male and female parts of the adapter could not be fully tightened due to a mismatch of the threads between the two pieces.It was reported that there was an issue with the connector.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARGYLE
Type of Device
CATHETER, PERITONEAL, LONG-TERM INDWELLING
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS   20101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key15031310
MDR Text Key303928780
Report Number3009211636-2022-00175
Device Sequence Number1
Product Code FJS
UDI-Device Identifier10884521004740
UDI-Public10884521004740
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K180485
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Pharmacist
Type of Report Initial,Followup,Followup,Followup
Report Date 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888415604
Device Catalogue Number8888415604
Device Lot Number2116900092
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-