Model Number 4410 |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported the device provided inaccurate etco2 values.No consequences or impact to patient were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
|
Event Description
|
The customer reported the device provided inaccurate etco2 values.No consequences or impact to patient were reported.
|
|
Manufacturer Narrative
|
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
|
|