ADVANCED BIONICS, LLC HIRES¿ 3D ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-04
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Purulent Discharge (1812); Unspecified Infection (1930)
Event Date 05/12/2022
Event Type  Injury  
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event Description
The recipient reportedly experienced an infection at the implant site.On may 12, 2022 the recipient was treated oral antibiotics.On (b)(6) 2022 the recipient underwent wound cleaning and antibiotics were administered intravenously.The purulent discharge persisted.The patient's device was explanted on (b)(6) 2022.
 
Manufacturer Narrative
The recipient's infection has reportedly resolved.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode array was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the electrical tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ 3D ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
alexandra gonzalez
28515 westinghouse place
valencia, CA 91355
MDR Report Key15045102
MDR Text Key296096070
Report Number3006556115-2022-01199
Device Sequence Number1
Product Code MCM
UDI-Device Identifier00840094455764
UDI-Public(01)00840094455764(11)211123(17)241031
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-