It was reported that shaft break occurred and the patient experienced bradycardia and arrhythmia.The 80% stenosed target lesion was located in the mildly tortuous and severely calcified left anterior descending artery.A 4.00 x 28mm synergy xd drug-eluting stent was selected for use.Upon loading on the wire, the wire exited the stent delivery before the monorail port.The wire was redirected and stent delivery was attempted.Upon inflation at about 1 to 2 atmospheres, a blood was noted in the stent delivery system and the balloon was leaking contrast.The delivery system was also noted to be broken at the proximal stent edge between the balloon and the monorail port.The device was not deployed and successfully retracted back into the guidezilla and everything was removed intact.The patient had arrhythmia and a neosynephrine was given for bradycardia.The procedure was completed with another of the same device.The patient was fully recovered.
|