Catalog Number 364314
Device Problem Inability to Auto-Fill (1044)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/19/2022
Event Type  malfunction  
Event Description
It was reported that bd preset¿ arterial blood collection syringe had insufficient blood flow through the device.The following information was provided by the initial reporter: "during using the abg it was found that the abg had insufficient blood flow issue.".
 
Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 5 retention samples from bd inventory were evaluated by visual examination and functional testing, each drawn with colored water, and no issues were observed relating to insufficient flow as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode insufficient blood flow.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRESET¿ ARTERIAL BLOOD COLLECTION SYRINGE
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15057717
MDR Text Key297986535
Report Number9617032-2022-00664
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K022426
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2023
Device Catalogue Number364314
Device Lot Number1197895
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/23/2022
Initial Date FDA Received07/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-