Catalog Number 5C4483
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the twist clamp of a minicap transfer set was "unable to close properly".This was observed during set up of the device for peritoneal dialysis therapy.The transfer set was replaced.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The event occurred on an unspecified date in (b)(6) 2022.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the sample was received for evaluation with a cap attached to the female connector and twist clamp in closed position.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing including leak, clear passage, and clamp function testing were performed with no issues noted.Torque engagement testing was performed and there were no issues.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15069741
MDR Text Key296489341
Report Number1416980-2022-03749
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007748
UDI-Public(01)00085412007748
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K192705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5C4483
Device Lot NumberH20L09029
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/28/2022
Initial Date FDA Received07/20/2022
Supplement Dates Manufacturer Received08/04/2022
Supplement Dates FDA Received08/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-