• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(M)
Device Problems Overheating of Device (1437); Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/30/2021
Event Type  malfunction  
Manufacturer Narrative
The customer notified cardioquip that a patient was involved during the malfunction, but there was no injury as a result, and no information about the patient was given.Cardioquip's investigation determined the heater to be the source of the issue.After all damaged components were replaced, the device passed final inspection, and is fully operational.
 
Event Description
The customer reported this device as overheating and smoking, with smoke damage being visible at the bottom of the tank.The user does not know how long the device was on the patient or what temperature the water was when the issue was noticed but they do know the setpoint was at 37 degrees.The patient was not injured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key15070488
MDR Text Key304614724
Report Number3007899424-2022-00079
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(M)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-