This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined he motor speeds were lower than specification; corroded housing caused the motor to become stuck.The shaft, reciprocating arm, housing, and various bearings were corroded; however, the device was not repairable.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to use of a device that was unrepaired and not within specifications ¿ usage error.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|