Model Number 10379675
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2022
Event Type  malfunction  
Manufacturer Narrative
Siemens has asked for the instrument to be returned for investigation.No one was harmed by this smoking analyzer and noproperty was damaged.The customer immediately unplugged the instrument when the smoke was seen.The customer is operational with a new instrument.The cause of this event is unknown.
 
Event Description
The customer reported that their clinitek status+ would not turn on and a burnt smell was observed.The customer attempted plugging the unit into a different outlet, and noted visible smoke.
 
Manufacturer Narrative
Siemens has completed the investigation.The d16 on the main board was source of the burning smell but it is not a fire risk.D16 fails when incorrect power is connected to the device, either through plugging in the wrong power supply (device uses a common barrel connector) or using 3.5v rechargeable batteries (device specifies using common 1.5v aa batteries).Additionally this can be caused naturally in the event of a power surge.Also worth noting the customer is unsure if the original power supply was in use at the time of this event.There is no definitive cause for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD
northern road
chilton industrial estates
sudbury, suffolk CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
thomas myers
2 edgewater drive
norwood, MA 02062
MDR Report Key15079181
MDR Text Key304808149
Report Number3002637618-2022-00048
Device Sequence Number1
Product Code JIL
UDI-Device Identifier00630414574639
UDI-Public00630414574639
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10379675
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/12/2022
Initial Date FDA Received07/21/2022
Supplement Dates Manufacturer Received09/02/2022
Supplement Dates FDA Received09/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-