Model Number 10379675 |
Device Problem
Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/12/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens has asked for the instrument to be returned for investigation.No one was harmed by this smoking analyzer and noproperty was damaged.The customer immediately unplugged the instrument when the smoke was seen.The customer is operational with a new instrument.The cause of this event is unknown.
|
|
Event Description
|
The customer reported that their clinitek status+ would not turn on and a burnt smell was observed.The customer attempted plugging the unit into a different outlet, and noted visible smoke.
|
|
Manufacturer Narrative
|
Siemens has completed the investigation.The d16 on the main board was source of the burning smell but it is not a fire risk.D16 fails when incorrect power is connected to the device, either through plugging in the wrong power supply (device uses a common barrel connector) or using 3.5v rechargeable batteries (device specifies using common 1.5v aa batteries).Additionally this can be caused naturally in the event of a power surge.Also worth noting the customer is unsure if the original power supply was in use at the time of this event.There is no definitive cause for this event.
|
|