• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problem Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2022
Event Type  malfunction  
Event Description
During implant, the stylet was unable to be removed from the left ventricular (lv) lead.The event was resolved by explanting and replacing the lv lead.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of stylet could not be removed was confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found that the connector pin and connector cap were pulled out of the connector assembly along with the inner coil which is consistent with procedural damage and the ptfe stylet coating was bunched up/clogged distal to the connector pin.The cause of the reported event was due to bunched up ptfe coating of the stylet that prevented the removal of the guidewire and excessive forces resulted in the connector pin and cap to be pulled out of the connector assembly.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15080947
MDR Text Key296419085
Report Number2017865-2022-15511
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000121004
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/05/2022
Initial Date FDA Received07/21/2022
Supplement Dates Manufacturer Received08/24/2022
Supplement Dates FDA Received09/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-