• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOQUIP, LLC CARDIOQUIP MODULAR COOLER HEATER; CARDIOPULMONARY BYPASS DEVICE Back to Search Results
Model Number MCH-1000(I)
Device Problems Fluid/Blood Leak (1250); Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2021
Event Type  malfunction  
Manufacturer Narrative
Cardioquip's investigation determined the water level sensor failed to activate causing the malfunction.After all damaged components were replaced, the device passed final inspection, and is fully operational.
 
Event Description
Customer complains of device smoking during cleaning procedure.
 
Event Description
Customer complains of device smoking during cleaning procedure.
 
Manufacturer Narrative
Further review determined that this is a non-reportable event.
 
Manufacturer Narrative
Further review determined that this is a non-reportable event.
 
Event Description
During their pm of the device, the technician had to fail the device due to smoke damage present which could not be resolved on-site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key15084393
MDR Text Key303638338
Report Number3007899424-2022-00083
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Inspection
Type of Report Initial,Followup,Followup
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(I)
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/22/2022
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received09/06/2022
09/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-