• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE NORMOFLO IRRIGATION FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE NORMOFLO IRRIGATION FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number H1100
Device Problem Mechanical Problem (1384)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device actuator pole does not come down.No patient injury was reported.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device was received with bent actuator pole and caster wheel.The customer stated problem was duplicated.Faulty actuator keypad switch was replaced.The cause of the reported problem could not be determined.A dhr (device history review) was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records., corrected data: d5: operator of the device: unknown e4: customer reported to fda: no information.H6: event problem patient code: 4582.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NORMOFLO IRRIGATION FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
attn: nanette / carlos
minneapolis, MN 55442
MDR Report Key15086333
MDR Text Key304084435
Report Number3012307300-2022-13983
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072080
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberH1100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-