• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Moisture or Humidity Problem (2986)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/12/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported the cardiosave intraaortic balloon pump (iabp) has a lot of condensation.Getting ce door not sealing error.It is unknown under which circumstances the event occurred or if patient was involved.
 
Event Description
It was reported the cardiosave intraaortic balloon pump (iabp) has a lot of condensation during routine check.Getting ce door not sealing error.There was no patient involvement.
 
Manufacturer Narrative
It was reported the cardiosave intraaortic balloon pump (iabp) has a lot of condensation during routine check.Getting ce door not sealing error.There was no patient involvement.A getinge field service engineer fse was dispatched to the site to evaluate the unit.He inspected unit for residual condensation per customers request and performed condensation removal , performed functional testing and safety check to factory specifications.Unit passes all functional and safety checks and is ready for patient use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15089588
MDR Text Key302025667
Report Number2249723-2022-01854
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.; UNKNOWN.
-
-