MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE; PISTON SYRINGE Back to Search Results
Model Number 305271
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that two bd integra¿ syringe with detachable needle experienced leakage past the stopper.The following information was provided by the initial reporter: multiple syringes when dispensing sample came out of the back of the syringe.Customer took 6 syringes and used same sample amount of acetone and 2 of the 6 syringes with little pressure sample came out of the back of the syringe.The sample was not viscous.
 
Manufacturer Narrative
H6: investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.A device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.
 
Event Description
It was reported that two bd integra¿ syringe with detachable needle experienced leakage past the stopper.The following information was provided by the initial reporter: multiple syringes when dispensing sample came out of the back of the syringe.Customer took 6 syringes and used same sample amount of acetone and 2 of the 6 syringes with little preassure sample came out of the back of the syringe.The sample was not viscous.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15089986
MDR Text Key297816083
Report Number1213809-2022-00421
Device Sequence Number1
Product Code MEG
UDI-Device Identifier30382903052715
UDI-Public30382903052715
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number305271
Device Catalogue Number305271
Device Lot Number0148581
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-