Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.Based on the received information from the field, the device did not provide sufficient benefit due to a partial migration of the active electrode out of cochlea, as confirmed by post-operative diagnostic imaging.Further, the recipient experienced facial nerve stimulation most likely due to electrical stimulation in the middle ear caused by the extra-cochlear channels.In addition, in situ measurements show three channels with hi and short circuit status due to undetermined reasons.This is a final report.
|