• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA500Q
Device Problems Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959); Noise, Audible (3273); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/06/2022
Event Type  malfunction  
Event Description
It was reported that the patient presented in clinic for a magnetic resonance imaging (mri) procedure while their implantable cardioverter defibrillator (icd) was not appropriately programmed with mri mode "on".The icd was noted to receive an auditory alert and then went into back-up operation with high voltage (hv) therapies disabled.The device could not be successfully interrogated upon going into back-up.The device was reprogrammed back to nominal settings and all functions were restored.The patient was discharged without any additional complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15094902
MDR Text Key304044924
Report Number2017865-2022-15963
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberP000123688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA; TENDRIL
Patient Age69 YR
Patient SexFemale
Patient Weight52 KG
-
-