Model Number N/A |
Device Problems
Electrical /Electronic Property Problem (1198); Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/18/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).South africa.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported by the surgeon that the dermatome was not getting enough power and that the speed is too slow.No adverse events were reported as a result of this malfunction.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was further reported that the unit was losing power.The issue was discovered prior to surgery.There was no patient involvement as it occurred prior to surgery and another device was used to complete the surgery.There was no surgical delay.The surgical technique was utilized.No adverse events were reported as a result of this malfunction.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d4; d9; g3; g6; h1; h2; h3; h4; h6.The event is confirmed.Review of the most recent repair record determined the speed slows and was unstable.The motor was replaced and resolved the reported issue.Review of complaint history found no additional related issues for this item and the reported part and lot combination.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|