STRYKER MEDICAL-KALAMAZOO BED INTOUCH WITH ZOOM MOTOR; BED, AC POWERED, ADJUSTABLE HOSPITAL
|
Back to Search Results |
|
Model Number 2141 |
Device Problem
Intermittent Loss of Power (4016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Event Description
|
This report summarizes 27 malfunction events, where it was reported the devices experienced the zoom suddenly stopping.There was 1 event with patient involvement; no adverse consequences were reported.
|
|
Manufacturer Narrative
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.27 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.There was no remedial action taken.This device is not labeled for single use.
|
|