• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESHAPE LIFESCIENCES LAP-BAND SYSTEM; ADJUSTABLE GASTRIC BAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESHAPE LIFESCIENCES LAP-BAND SYSTEM; ADJUSTABLE GASTRIC BAND Back to Search Results
Model Number C-2365
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Pain (1994)
Event Date 06/21/2022
Event Type  malfunction  
Event Description
Surgeon reported a port tubing dislocation in a patient that he implanted a new band system into one week ago.Patient lap band placed (b)(6) 2022 without issue, then 5 days later had pelvic pain and a ct at er showed disconnected tubing hanging in pelvis.
 
Manufacturer Narrative
The observed failure appears to be consistent with an overlarge down or lateral application of mechanical stress to the tube at or near the external collar of the port housing.While it is impossible to know how the mechanical stress was exerted onto the access port, we do know that variations in port placement can be an issue.The instructions and illustration of optimal port placement are included in the lap-band systems instructions for use (ifu): failure to create a stable, smooth path for the access port tubing, without sharp turns or bends, can result in tubing breaks and leakage.In order to avoid incorrect placement, the port should be placed lateral to the trocar opening.A pocket must be created for the port so that it is placed far enough from the trocar path to avoid abrupt kinking of the tubing.The tubing path should point in the direction of the access port connector so that the tubing will form a straight line with a gentle arching transition into the abdomen as shown in the diagrams below.No new risks identified, the current risk for this event is with a low rate of occurrence for the reported complaint category.No correction or corrective action required.The lot history record for the port lot and band was reviewed and no discrepencies or non-conformances recorded.Product was manufactured according to specification.Unable to determine root cause.The investigation findings do not lead to a clear conclusion about the cause of the reported adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
RESHAPE LIFESCIENCES
1001 calle amanecer
san clemente CA 92673
Manufacturer Contact
maria quiroz
1001 calle amanecer
san clemente, CA 92673
8449377374
MDR Report Key15108974
MDR Text Key304611728
Report Number3013508647-2022-00062
Device Sequence Number1
Product Code LTI
UDI-Device Identifier00850007764654
UDI-Public(01)00850007764654(17)231001(10)75HA6562(11)211001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/01/2023
Device Model NumberC-2365
Device Catalogue NumberC-2365
Device Lot Number75HA6562
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/28/2022
Initial Date FDA Received07/26/2022
Supplement Dates Manufacturer Received06/28/2022
Supplement Dates FDA Received09/01/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-