• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME, PNEUMATICALLY-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME, PNEUMATICALLY-POWERED Back to Search Results
Model Number N/A
Device Problems Power Problem (3010); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record identified the following related repair: tech confirmed the reported issue as the motor had seized during assessment.Additionally the tech found that the machined head was damaged and screws were worn, the unit was out of calibration, and the control bar was out of position.Tech replaced the motor, motor sleeve, machined head, shaft bearings, sleeve bearings, screws, and then repositioned the control bar.Tech tested the unit and returned it to service.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the handpiece gets stuck.During initial evaluation of the device, the motor exhibited inconsistent speed including seizing.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
Following sections were updated or corrected : g3 if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME, PNEUMATICALLY-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15111811
MDR Text Key302187594
Report Number0001526350-2022-00704
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number60223471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2022
Was the Report Sent to FDA? No
Date Manufacturer Received07/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-