• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG SOLUTIONS S.A. DIALYSIS UNKNOWN; CATHETER, PERITONEAL, LONG-TERM INDWELLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG SOLUTIONS S.A. DIALYSIS UNKNOWN; CATHETER, PERITONEAL, LONG-TERM INDWELLING Back to Search Results
Model Number DIALYSIS UNKNOWN
Device Problem Break (1069)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
According to the reporter, the catheter's red or blue end had developed a split/crack (back ends).There was no available repair kit and the catheter remained in use.There was no reported patient outcome.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, after 9 months, 1 week and 5 days, the catheter's red (arterial) and blue (venous) luer adapters (back end) had developed a split/crack and it was documented on july 22, 2022.There was no leak.Tego was not utilized.Clinell 2% chlorhexidine in 70% alcohol individually wrapped wipes, unisurge pack line on (3ml and 5ml syringes), gambro artiset blood tubing system for hemodialysis with gambro artis, unisurge pack line off (3ml and 5ml syringes) with red uhs obturator cap, bd plastipak 20ml luer-lok syringe, bd vacutainer multiple sample luer adapter, bd vacutainer one-use holder, bd vacutainer blood collection tubes were the products utilized with the device.Clinell 2% chlorhexidine in 70% alcohol individually wrapped wipe was the cleaning agent used onthe device and it was typically utilized to clean the adapters.The adapters were tighten by hand.Nothing unusual documented at time of insertion.Flushing at time of insertion was not documented.The catheter had heparin lock withdrawn and lumen flushed prior to every use.No excessive force used on the catheter.There was no available repair kit as it was out of stock.The manufacturer was contacted for advice on next steps, and datix (safety report) was completed.Quality and safety team of the hospital investigated, and dated photographs taken of both catheter lumens.New monitoring assessment tool was made to carryout daily assessment of lumens and catheter.Renal team researched different brand of hemodialysis catheter to replace the catheter.The dialysis catheter would need to be removed and replaced if appropriate due to the risk of infection and air embolus.The dialysis catheter was removed.There was no blood loss.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIALYSIS UNKNOWN
Type of Device
CATHETER, PERITONEAL, LONG-TERM INDWELLING
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS  20101
Manufacturer (Section G)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
CS   20101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view,mn, CO 55112
7635265677
MDR Report Key15112904
MDR Text Key303577255
Report Number3009211636-2022-00186
Device Sequence Number1
Product Code FJS
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIALYSIS UNKNOWN
Device Catalogue NumberDIALYSIS UNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
Patient Weight47 KG
-
-