Model Number N/A
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2022
Event Type  malfunction  
Event Description
It was reported that during set up prior to surgery dermatome handpiece was hard to connect to hose, it required more force to lock in place then should be necessary.Another device was used and the hose connected properly.No harm or delay were reported.No other adverse events were reported as it relates to this event.
 
Manufacturer Narrative
This incident has been recorded under (b)(4).Once investigation is completed a final/supplemental report will be submitted.
 
Manufacturer Narrative
This incident has been recorded under (b)(4).Review of the most recent repair record determined that the hose was hard to connect.The o-ring on the swivel and hose were replaced and resolved the reported issue a definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15115218
MDR Text Key298692539
Report Number0001526350-2022-00713
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number62554522
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received09/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-