• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number CAX500T12
Device Problem Degraded (1153)
Patient Problems Dyspnea (1816); Renal Failure (2041); Sore Throat (2396); Unspecified Respiratory Problem (4464)
Event Date 08/17/2021
Event Type  Injury  
Event Description
The manufacturer received information alleging a cpap device's sound abatement foam became degraded and caused nasal/throat soreness and irritation, difficulty breathing, lungs were ingested and kidney failure.The patient did not report to receive medical intervention.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was mention of dirt/dust contamination around the air inlet of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There were no errors found.The manufacturer concludes that they could not confirm the customer's allegation and there was no visible foam degradation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
PHILIPS RESPIRONICS
6501 living place
pittsburgh PA 15206
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key15116771
MDR Text Key296748044
Report Number2518422-2022-66362
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberCAX500T12
Device Catalogue NumberCAX500T12
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
Patient Outcome(s) Other;
-
-