• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HEAD 42MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74121142
Device Problems Biocompatibility (2886); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Metal Related Pathology (4530); Insufficient Information (4580)
Event Date 03/29/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that after a birmingham hip resurfacing procedure on (b)(6) 2006 due to osteoarthritis in the right hip, the patient underwent a revision surgery to a total hip arthroplasty on (b)(6) 2022.The resurfacing femoral head was explanted.The current health status of the patient remains unknown.
 
Event Description
It was reported that after a birmingham hip resurfacing procedure on (b)(6) 2006 due to the patient's osteoarthritis in the right hip, the patient underwent revision surgery to a total hip arthroplasty on (b)(6) 2022 due to pain and elevated levels of cobalt.During surgery, fluid was seen in the pseudotumor.The acetabular cup and the femoral head were explanted.It is unknown the current status of the patient.
 
Manufacturer Narrative
H3, h6: it was reported that a right hip revision surgery was performed due to pain and elevated levels of cobalt.During surgery, fluid was seen in the pseudotumor.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the devices concerned was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for acetabular cup and femoral head.However, as the devices are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.The clinical information provided, of the elevated cobalt, and pseudotumor may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 42MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15121155
MDR Text Key296765936
Report Number3005975929-2022-00438
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502773
UDI-Public03596010502773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2011
Device Model Number74121142
Device Catalogue Number74121142
Device Lot Number66429
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/15/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/15/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2745-2015 / Z-2747-20
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age64 YR
Patient SexMale
-
-