MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST TRADITION PRO TB; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST TRADITION PRO TB; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 790345
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
This report summarizes 1 malfunction event where a midwest tradition pro handpiece would not hold burs.No injury resulted in the event.
 
Manufacturer Narrative
1 of 1 device was returned for evaluation.Evaluation of 1 device found normal chuck wear and the turbine needed to be replaced.The device was repaired and returned to the customer.This event is being submitted as part of vmsr.Only one event was received for this device during this quarter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST TRADITION PRO TB
Type of Device
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494593
MDR Report Key15124586
MDR Text Key304463737
Report Number9614977-2022-00045
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
PMA/PMN Number
K131319
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number790345
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/28/2022
Patient Sequence Number1
-
-