Catalog Number 254041
Device Problem Device Ingredient or Reagent Problem (2910)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used as a default a device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that while using bd plate sabouraud ag w/gent w/chmp 90mm 20 atypical growth was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "herewith to report the abundant growth of gram-positive and gram-negative bacteria on various soils in sabouraud.".
 
Event Description
It was reported that while using bd plate sabouraud ag w/gent w/chmp 90mm 20 atypical growth was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "herewith to report the abundant growth of gram-positive and gram-negative bacteria on various soils in sabouraud".
 
Manufacturer Narrative
H.6.Investigation summary event description: it was reported that several plates would not be selective.Complaint history review: the complaint history has been reviewed and for this product, several complaints were reported for a performance issue.Therefore, a trend was identified, and further investigation was performed.Batch history record (bhr) review: the batch history has been reviewed.It was identified that the chloramphenicol containing in this product was probably affected and further finished products were identified.Sample analysis: a performance test has been conducted on the retain samples of the complained batch 2150314.The results showed not deviation; however while testing other batches containing chloramphenicol, the performance issue could be confirmed.Return samples were not provided; however, a picture was shared showing growth on plates.Evaluation results and investigation conclusion: based upon our investigation, we have excluded any systematic failure in our manufacturing process.All the release testing was satisfactory and also the retest performed on the retain samples of the complained batch 2150314 did not reveal any deviation to the requirement.However while testing other batches containing chloramphenicol, the performance issue could be confirmed for excessive growth.Bd has identified an outside supplier issue with this plate¿s impregnated antibiotic, chloramphenicol, which results in a reduction in the antibacterial activity of the chloramphenicol over the shelf life of the agar.As a result a product recall has been initiated for the affected lots in the affected regions.Based on the above mentioned evaluation, this complaint was confirmed.Bd regrets the inconveniences you have been experiencing due to this issue and bd will continue monitoring incoming complaint with similar failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PLATE SABOURAUD AG W/GENT W/CHMP 90MM 20
Type of Device
CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON DICKINSON GMBH
8-12 tullastrasse
heidelberg 69126
GM  69126
Manufacturer (Section G)
BECTON DICKINSON GMBH
8-12 tullastrasse
heidelberg 69126
GM   69126
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15124931
MDR Text Key304335072
Report Number9680577-2022-00086
Device Sequence Number1
Product Code JSJ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/01/2022
Device Catalogue Number254041
Device Lot Number2150314
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-