• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR WITH KANAMYCIN AND VANCOMYCIN (KV); CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR WITH KANAMYCIN AND VANCOMYCIN (KV); CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL Back to Search Results
Model Number 221736
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2022
Event Type  malfunction  
Event Description
It was reported that while using 100 bd bbl cdc anaerobe 5% sheep blood agar with kanamycin and vancomycin (kv) bacterial contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "bacterial (based on photo review)".
 
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Manufacturer Narrative
H.6 investigation summary: during manufacturing of material 221736, media is formulated and sent through a high temperature short time sterilizer to remove bioburden.The petri dishes are subjected to uv radiation to decrease bioburden.The petri dishes are filled in a positive pressure hepa filtered environment.The filled plates are cooled and immediately wrapped into sleeves to decrease the introduction of microbes.Sleeves are then packaged into cartons and then transferred to a refrigerated truck (2 to 8 degrees c) for shipment to the distributor.Bd distributors are provided with the storage guidelines for the shipping and handling of bd media of 2 to 8 degrees c in a dark place.The batch history record for batch 2118857 was satisfactory and no quality notifications were generated during manufacturing and inspection.The release testing that is performed on this product does include physical attribute and bioburden testing.A sample of plates are incubated at 25 degrees c and at 35 degrees c for approximately 72 hours.All physical attribute and bioburden testing performed on this batch was satisfactory per bd internal procedures.Affected product does not have any sterility claims; the product is tested for bioburden prior to release to ensure that it conforms to product specifications.However, this does not ensure that the end-user will not receive a contaminated plate.The complaint history was reviewed, and no other complaints have been taken on batch 2118857.Retention samples from batch 2118857 were not available for inspection.Two photos were received for investigation.One photo shows a sleeve from batch 2118857 with at least two broken plates within the sleeve.The second photo shows the top of an unopened sleeve with what appears to be a colony of growth visible in the top plate within the sleeve.The top plate is also broken.No return samples were received for investigation.Review of the manufacturing record for this batch did not find a cause for broken plates from the production.This complaint can be confirmed for broken plates and contamination.This product does not have a sterile claim and no other complaints have been taken on this batch for contamination.No complaint trends for broken plates or contamination have been identified for this product; no actions are indicated at this time.Bd will continue to trend complaints for broken plates and contamination.
 
Event Description
It was reported that while using 100 bd bbl¿ cdc anaerobe 5% sheep blood agar with kanamycin and vancomycin (kv) bacterial contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "bacterial (based on photo review)".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ CDC ANAEROBE 5% SHEEP BLOOD AGAR WITH KANAMYCIN AND VANCOMYCIN (KV)
Type of Device
CULTURE MEDIA, SELECTIVE AND NON-DIFFERENTIAL
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15135834
MDR Text Key304448559
Report Number1119779-2022-01045
Device Sequence Number1
Product Code JSJ
UDI-Device Identifier10382902217364
UDI-Public10382902217364
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K803022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/15/2022
Device Model Number221736
Device Catalogue Number221736
Device Lot Number2118857
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2022
Initial Date FDA Received07/29/2022
Supplement Dates Manufacturer Received10/18/2022
Supplement Dates FDA Received10/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-