It was reported that the patient lost therapy and the ipg did not communicate with external devices following an unrelated procedure.Patient does not recall if the ipg was set into surgery mode prior to the procedure.Troubleshooting was unable to resolve the issue.As such, surgical intervention took place on (b)(6) 2022 wherein the ipg was explanted and replaced with a new ipg addressing the issue.Reportedly, therapy was confirmed post op.
|
Corrected data: h6 codes.The reported observation of ¿ipg - no communication¿ was confirmed.The analysis results concluded the inability to establish communication between the programmer and the implantable pulse generator (ipg) was due to the device entering the service application state.The root cause of the device entering the service application state was due to the patient's dental procedure as the device stopped communicating after that.Per the clinicians manual: per clinicians manual arten600149416 - under warnings - there is sufficient documentation concerning the use of use short-wave diathermy, microwave diathermy, or therapeutic ultrasound diathermy (all now referred to as diathermy) on patients implanted with a neurostimulation system, and medical procedures (such as therapeutic radiation and electromagnetic lithotripsy).Per the clinicians manual electrosurgery.To avoid harming the patient or damaging the neurostimulation system, do not use monopolar electrosurgery devices on patients with implanted neurostimulation systems.Before using an electrosurgery device, place the device in surgery mode using the patient controller app or clinician programmer app.Confirm the neurostimulation system is functioning correctly after the procedure.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|