• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE; DERMATOME HANDPIECE, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE; DERMATOME HANDPIECE, PNEUMATIC Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Patient Device Interaction Problem (4001)
Patient Problem Abrasion (1689)
Event Date 07/13/2022
Event Type  Injury  
Event Description
It was reported that at the start of use the black catch that moves the knives back and forth was loose, and the dermatome was not functioning properly.A new leg skin had to be taken as the initial graft was damaged.A few minute delay occurred, and another device was used to complete procedure.No additional patient consequences were reported as a result of this malfunction.
 
Manufacturer Narrative
This incident has been recorded under (b)(4).Once investigation is completed a final/supplemental report will be submitted.
 
Manufacturer Narrative
The results of the investigation are as follows: review of the most recent repair record determined the blade movement was loose due to a broken neck pin and damaged back cover.Additionally, all the neck screws were broken off and stuck into the housing.The neck and neck screws were replaced.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15140936
MDR Text Key297002237
Report Number0001526350-2022-00728
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)170504(10)63618116
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number63618116
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-