Model Number N/A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Visual examination of the returned device found it to exhibit signs of repeated use and have one of components 42527991001 and 42527991002 disassembled / missing.The components were not returned.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-02169, 0001822565-2022-02170, 0001822565-2022-02171, 0001822565-2022-02172, 0001822565-2022-02173.
 
Event Description
It was reported that the instrument is missing ball bearings.There was no known patient involvement.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE GH 10 MM THICKNESS
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15142351
MDR Text Key297076275
Report Number0001822565-2022-02167
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024245945
UDI-Public(01)00889024245945(10)63863382
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42527900700
Device Lot Number63863382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2022
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-