MAUDE Adverse Event Report: RESPIRONICS, INC DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX200H11
Device Problem Degraded (1153)
Patient Problems Ulcer (2274); Sore Throat (2396); Respiratory Tract Infection (2420)
Event Date 09/24/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging sores in her nose, nasal/throat irritation or soreness related to a cpap device's sound abatement foam.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted where the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging visualization of particles related to a cpap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Section h6 updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC
1001 muury ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 muury ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburg, PA 15206
2673970028
MDR Report Key15142437
MDR Text Key301128105
Report Number2518422-2022-67357
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX200H11
Device Catalogue NumberDSX200H11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
Patient SexFemale
-
-