ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problems Failure to Form Staple (2579); Failure to Fire (2610); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2022
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure the customer reports that the device was not working properly.The procedure was successfully completed with no patient consequences.
 
Manufacturer Narrative
(b)(4).Batch # x9474n.Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device lot number, and no [non-conformances / manufacturing irregularities] were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.Additional information was requested, and the following was obtained: please clarify how the ¿the clipper is not working properly¿.The device locked after second shot.Did device not feed clips? sometimes.Did device feed clips sideways? sometimes.Did device not fire clips (jammed)? yes.Did device fire malformed clips? yes.Did device fire scissored clips? yes.Did device drop or eject clips? yes.The first two if other, specify." this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15152948
MDR Text Key304874723
Report Number3005075853-2022-05059
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberX9474N
Was Device Available for Evaluation? No
Date Manufacturer Received07/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-