Model Number CDHFA500Q
Device Problem No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/15/2022
Event Type  malfunction  
Event Description
It was reported the patient presented for an implantable cardioverter defibrillator (icd) upgrade.During implant, the icd exhibited a loss of pacing output.The physician elected to complete the procedure with a different icd.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of loss of pacing output could not be confirmed.The implantable cardioverter defibrillator (icd) was returned for analysis.Telemetry, impedance, sensing, pacing, and high voltage output functions were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15153753
MDR Text Key302426410
Report Number2017865-2022-16650
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberS000086031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUARTET; TENDRIL; TENDRIL STS
Patient Age45 YR
Patient SexMale
-
-