This event has been recorded by zimmer biomet under (b)(4).(b)(6).Investigation is complete.This is an initial final report submission.Review of the most recent repair record identified the following related repair: the screw was worn, the motor speed was inconsistent, and the cable was damaged; however, no wires appear to be exposed.The screw, motor, and plug harness assembly were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|