It was reported that a proximal filter dislodgement occurred.A sentinel cerebral protection system (cps) was used during a transcatheter aortic valve implantation (tavi) procedure.An introducer sheath was placed into the radial artery and the sentinel cps was advanced and positioned with the proximal filter into brachiocephalic artery.The proximal filter was successfully deployed into the brachiocephalic artery.Upon cannulating to the left common carotid artery with the articulating distal sheath, the proximal filter dislodged from the brachiocephalic artery in an open state.The proximal filter was recaptured and repositioned in the brachiocephalic artery under fluoroscopy.However, the proximal filter slider exhibited resistance and the proximal filter would not open.The sentinel cps was successfully retrieved and replaced with another sentinel cps to complete the procedure.No patient complications were reported.
|
Device evaluated by mfr.: the sentinel cps was returned to boston scientific (bsc) and was analyzed by a bsc quality engineer.Visual analysis revealed a sheathed proximal filter, relaxed articulating distal sheath, unsheathed distal filter, and outer shaft kink near the handle.X ray inspection revealed no damage in the braids of the inner shaft.A functional test was performed and found that before flushing the sentinel cps, the proximal filter was unable to be unsheathed using the proximal filter slider (#1).After flushing the sentinel cps, the proximal filter was still unable to be unsheathed using the proximal filter slider (#1).A dissection test was performed using scissors to create a transversal cut on the distal part of the outer shaft.After the dissection, no damage was observed in the braids of the inner shaft.
|
It was reported that a proximal filter dislodgement occurred.A sentinel cerebral protection system (cps) was used during a transcatheter aortic valve implantation (tavi) procedure.An introducer sheath was placed into the radial artery and the sentinel cps was advanced and positioned with the proximal filter into brachiocephalic artery.The proximal filter was successfully deployed into the brachiocephalic artery.Upon cannulating to the left common carotid artery with the articulating distal sheath, the proximal filter dislodged from the brachiocephalic artery in an open state.The proximal filter was recaptured and repositioned in the brachiocephalic artery under fluoroscopy.However, the proximal filter slider exhibited resistance and the proximal filter would not open.The sentinel cps was successfully retrieved and replaced with another sentinel cps to complete the procedure.No patient complications were reported.
|