Model Number G247 |
Device Problems
Pacing Problem (1439); Failure to Read Input Signal (1581); Unexpected Therapeutic Results (1631); Under-Sensing (1661)
|
Patient Problems
Cardiac Arrest (1762); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/30/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) undersensed the patients ventricular fibrillation (vf) which led to delay in therapy.Upon further review, it was noted there was some inappropriate ventricular pacing being delivered during the ventricular fibrillation (vf) episode.Additionally, the vf was quite fine.Currently, the patient is hospitalized on extracorporeal membrane oxygenation (ecmo).At this time, the device remains in service.
|
|
Manufacturer Narrative
|
Correction to b2, and h6 fields (patient code and impact code).
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) undersensed the patients ventricular fibrillation (vf) which led to delay in therapy.Upon further review, it was noted there was some inappropriate ventricular pacing being delivered during the ventricular fibrillation (vf) episode.Additionally, the vf was quite fine.Currently, the patient is hospitalized on extracorporeal membrane oxygenation (ecmo).At this time, the device remains in service.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) undersensed the patients ventricular fibrillation (vf) which led to delay in therapy.Upon further review, it was noted there was some inappropriate ventricular pacing being delivered during the ventricular fibrillation (vf) episode.Additionally, the vf was quite fine.Currently, the patient is hospitalized on extracorporeal membrane oxygenation (ecmo).At this time, the device remains in service.
|
|
Manufacturer Narrative
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.Correction to b2, and h6 fields (patient code and impact code).
|
|
Search Alerts/Recalls
|
|