• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER Back to Search Results
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/22/2022
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the survey respondent stated the bardex 3-way ptfe coated latex foley catheter, 30 ml balloon, did not perform as intended and elaborated as hematuria not clearing as much when asked if the foley catheter performed as intended and resulted in the clinical outcome of successful urinary drainage from the bladder or from the nephrostomy placement.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿walls do not have enough hoop strength¿.It was unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Do not resterilize single use only.Do not reuse.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Consult instructions for use.Do not use if package is damaged.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulation.¿ correction: e.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the survey respondent stated the bardex 3-way ptfe coated latex foley catheter, 30 ml balloon, did not perform as intended and elaborated as hematuria not clearing as much when asked if the foley catheter performed as intended and resulted in the clinical outcome of successful urinary drainage from the bladder or from the nephrostomy placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATEX FOLEY CATHETER
Type of Device
LATEX FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15174896
MDR Text Key302606613
Report Number1018233-2022-06092
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/22/2022
Initial Date FDA Received08/05/2022
Supplement Dates Manufacturer Received09/08/2022
Supplement Dates FDA Received09/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-