The mayfield modified skull clamp (a1059) was returned for evaluation: failure analysis: evaluation confirmed that the unit needed repair.Unit received with the lock having both rotational and lateral movement and a residue buildup was present.Upon disassembly, repair noted the index knob and the lock needed new components added to replace worn internal parts.Unit was machined to have heli-coils added to the large starburst threads.General cleaning performed and worn internal parts were replaced.Unit was sent to quality engineering for further investigation due to a skull fracture post placement and further investigation by quality engineering confirms the findings from the service & repair evaluation, no additional device deficiencies found.Root cause: the unit passed all specific functional testing requirements, except for the lock having rotational movement due to routine use and wear - unit is beyond integra¿s 7 years recommended life cycle (manufactured 2011).When the unit is properly positioned and put under pressure the unit will function properly.Probable root cause for the reported complaint is improper setup of the skull clamp on the patient.No corrective or preventive action is required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|