Model Number EMAX2PLUS |
Device Problems
Material Frayed (1262); Unintended System Motion (1430)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be traced to user, which is user error.Udi: (b)(4).
|
|
Event Description
|
It was reported by china that during service and evaluation, it was determined that the motor device had a damaged wire, unintended activation/motion, unexpected noise, could not secure/lock cutter and had cord damage.It was further determined that the device failed pretest for visual assessment: (verify that the cord is not damaged), cutter lock assessment, motor thermistor assessment, safety assessment: (cutter does not rotate, ¿e2¿ is displayed), air pump assessment and hand control assessment.It was noted in the service order that during an unspecified surgical procedure, it was discovered that the cord was damaged.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was available to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|