Model Number TSH
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2022
Event Type  malfunction  
Manufacturer Narrative
The sample from the patient was submitted for investigation where the customer's high tsh value was confirmed.Upon further analysis, no interfering factors for the tsh assay were detected.Using size exclusion chromatography, a macrotsh was not present in the sample.Differences between the tsh results generated with the different analyzers are caused by differences in the setups of the assays, the antibodies used, and differences in the standardization materials and procedures used.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination, and other findings.From the information provided and the analysis of the sample, a general reagent issue could be excluded.
 
Event Description
The initial reporter received questionable elecsys tsh assay results on a cobas e 801 module.On (b)(6) 2022, the initial tsh result was 22.4 miu/l.The result was reported outside of the laboratory.The physician questioned the result as it did not fit the patient's clinical picture.On (b)(6) 2022, the sample was repeated on an abbott architect analyzer.The tsh result was 13.7802 uui/ml.The sample was also repeated on the e801 with a peg dilution.The tsh result was 7.86 miu/l.The peg result does not match the clinical results either; for this reason, the physician suspects a possible interference (e.G.A macro-tsh).It is not known which result was deemed correct.The reagent expiration date was requested but not provided.This medwatch covers tsh.Refer to the medwatch with patient identifier (b)(6) for information on the ft4 iii results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15184617
MDR Text Key302927560
Report Number1823260-2022-02358
Device Sequence Number1
Product Code JLW
UDI-Device Identifier07613336157633
UDI-Public07613336157633
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTSH
Device Catalogue Number08443432190
Device Lot Number622589
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
AMIODARONE; EUTIROX 75; FUROSEMIDE
Patient Age52 YR
Patient SexFemale
-
-