• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Catalog Number UE514010300
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2022
Event Type  malfunction  
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the cart is reading full when its actually empty.The event timing was outside of surgery.There was no harm or delay reported.No adverse events were reported as a result of this malfunction.
 
Event Description
There is no additional event information available.
 
Manufacturer Narrative
The unit was reading cylinder 1 as full when it was empty.The level sensor was replaced.Dhr was reviewed and no discrepancies related to the reported event were found.The level sensor is required to read and report fluid level in the cylinders in order for the unit to function as intended.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR - REFURBISHED
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15190614
MDR Text Key304921931
Report Number0001954182-2022-00073
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024501645
UDI-Public(01)00889024501645
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUE514010300
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-