Catalog Number 5C4483
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/17/2022
Event Type  malfunction  
Event Description
It was reported that a minicap transfer set was ¿unable to close properly¿.This was observed during setup of the device for peritoneal dialysis therapy.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to h3, h6 and h10.H10: the sample was received for evaluation.A visual inspection with naked eye was performed with no issues noted.Functional testing including leak, clear passage, clamp function, and torque engagement testing were performed with no issues noted.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15193297
MDR Text Key300908525
Report Number1416980-2022-04138
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007748
UDI-Public(01)00085412007748
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K192705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4483
Device Lot NumberH20L09029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/19/2022
Initial Date FDA Received08/09/2022
Supplement Dates Manufacturer Received08/22/2022
Supplement Dates FDA Received09/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-