Model Number IPN000072 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that "the guide wire gets stuck at the tip of the catheter.It happened with two catheters.Third one worked properly".No patient harm, injury, or consequence reported.Patient condition reported as "fine".Additional information received stated that the two catheters were used on the same patient, part and lot number were the same for both devices.No issues with the third catheter.Associated to mdr 3010532612-2022-00310.
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint that the "guide wire gets stuck at the tip of the catheter" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.If the product is returned at a later date, a full investigation of the sample will be completed.
|
|
Event Description
|
It was reported that "the guide wire gets stuck at the tip of the catheter.It happened with two catheters.Third one worked properly".No patient harm, injury, or consequence reported.Patient condition reported as "fine".Additional information received stated that the two catheters were used on the same patient, part and lot number were the same for both devices.No issues with the third catheter.Associated to mdr 3010532612-2022-00310.
|
|
Manufacturer Narrative
|
(b)(4).The reported lot number matches the lot number on the returned original packaging pouch.Returned for investigation was a 6fr.Wedge 110cm catheter with its original packaging pouch.The sample was returned in a cardboard box and was in a sealed bio-hazard bag.The sample was loosely packed within the original packaging pouch.Upon return, the supplied 1.0cc control stroke syringe was connected to the inflation lumen stopcock; no damage or abnormalities were noted to the returned syringe.The inflation lumen stopcock was in the open position.The recommended volume capacity of the balloon is 1.0cc.Upon microscopic inspection, a wrinkle was noted on the balloon surface.No condensation was noted in the inflation lumen extension line.No contrast media was noted within the injection lumen extension line.No blood was noted on the interior or the exterior surfaces of the returned sample.Upon further inspection, no blockage or abnormalities were noted to the distal tip of the catheter.No kink, bend or flattened surface was noted on the returned catheter extrusion/body.No other damage or abnormalities were noted to the returned sample.The inflation lumen was injected with 1.0cc of air using the returned control stroke syringe.The balloon inflated symmetrically.One side of the balloon measured approximately 5mm.The other side measured approximately 5mm.The balloon did meet specifications per manufacturing graphic of radius ratio less than or equal to 1.5.The inflation lumen was injected with 1.0cc of air using the returned control stroke syringe.The balloon inflated symmetrically.The balloon deflated in less than 3 seconds when the syringe was removed per internal manufacturing specification.No pull away was noted after the tug test.The balloon was placed in water and air was injected into the inflation lumen again.No leak was noted.A lab inventory 0.025in guidewire was back loaded through the distal tip of the catheter.No resistance was noted; the guidewire was able to advance through without any difficulty.No blood or debris was noted.The guidewire was front loaded through the injection extension luer.No resistance was noted; the guidewire was able to advance through without any difficulty.No blood or debris was noted.The injection lumen was aspirated and flushed.No blood or debris was noted.The reported complaint that the "guide wire gets stuck at the tip of the catheter" is not confirmed.A device history record review was performed, and no relevant findings were identified.The root cause of the complaint is undetermined.Teleflex will continue to monitor and trend for reports of this nature.
|
|
Event Description
|
It was reported that "the guide wire gets stuck at the tip of the catheter.It happened with two catheters.Third one worked properly".No patient harm, injury, or consequence reported.Patient condition reported as "fine".Additional information received stated that the two catheters were used on the same patient, part and lot number were the same for both devices.No issues with the third catheter.Associated to mdr 3010532612-2022-00310.
|
|
Search Alerts/Recalls
|