Catalog Number 3325040 |
Device Problem
Device Damaged Prior to Use (2284)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/26/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that when the box was opened, the ampule was broken - seal was still intact.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: then device associated with this complaint was received for examination.Visual examination found the ampule broken with the seal still in the ampule package.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|